设为首页加入收藏
轮播广告
产品详情
Name:CHI860D electrochemical workstation
The equipment used in electrochemical synthesis basic research, basic research and teaching impedance analysis, battery materials development, bio-electrochemical (sensors) and metallurgy, pharmaceuticals, environmental analysis and other areas of analysis.
The equipment used in electrochemical synthesis basic research, basic research and teaching impedance analysis, battery materials development, bio-electrochemical (sensors) and metallurgy, pharmaceuticals, environmental analysis and other areas of analysis. 

The meter can work in two, three, or four electrode pattern. Four-electrode mode can be used liquid / liquid interface electrochemical measurements or high current / low impedance electrolytic cell (such as batteries) system, using a four-electrode can effectively eliminate measurement errors due to cable and contact resistance caused. There are instruments external signal input channel, the external input voltage signal can be recorded in the record, while the electrochemical signal, such as spectral signals. This spectrum is very convenient electrochemical experiments. 

First, the performance characteristics 

(1) Software Features 
Test system software interface, easy operation and intuitive. Real-time observation window, the use of graphical data integration window, making the testing process intuitive and efficient. Software running on the users operation and data validity and completeness checks carried out adequately and timely prompt or warning. 
(2) real-time data display has a data list function, the impedance measurement results can be displayed in real time and open the data file, can easily view the results. 
(3) be able to save the data in real-time result, even if a sudden power failure without losing data, improve your data safe. 

Hardware parameter index 
Potential range: ± 10V 
Potential rise time: <1 microseconds 
Tank pressure: ± 26V 
Electrode system: two, three or four electrodes electrode 
Current range: ± 2A 
Reference electrode input impedance: 11013 ohm 
Sensitivity: 110-12-0.1A / V 12 ranges 
Input bias current: <50pA 
Current measurement resolution: <0.01pA 
Minimum potential increment in CV: 0.1mV 
Potential update rate: 10MHz 
Fast data collection: 16 @ 1MHz resolution 
External voltage input signal recording channels 
Automatic and manual iR drop compensation 
CV and LSV scan speed: 0.000001-5,000V / s 
When the potential sweep potential increment: 0.1mV@1000V/s 
CA and CC Pulse Width: 0.0001-1,000sec 
CA and CC step Views: 320 
DPV and NPV pulse width: 0.0001-10sec 
SWV Frequency: 1-100kHz 
ACV Frequency: 0.1-10kHz 
SHACV Frequency: 0.1-5kHz 
IMP Frequency: 0.00001-1MHz 
Automatic potential and current zeroing 
Potential and current measuring low-pass filter, automatic or manual settings, 
Covering the frequency range of eight orders of magnitude 
Control output cell: nitrogen, with stirring, percussion 
Maximum data length: 128,000 points -4,096,000 point selectable 
Instrument Size: 32cm (W) 28cm (deep) 12cm (high) 
Instrument Weight: 7.5kg
 
脚注信息